Digital output 1
You have the following setting options in this menu:
menu item
4.2
S
Function
, menu item
4.2.1
You can configure digital output 1 as a pulse, frequency, alarm or
status output (see menu Table 5-1)
S
Signal type
, menu item
4.2.2
You can configure different signal types for the output signal of
digital output 1:
-
Active:
The device-internal voltage is used (+24 V).
-
Passive:
External supply required.
U
Active
24 V
4 7 8 3 5
U
Passive
4 7 8 3 5
External supply
max. DC 30 V
I
Counter
Counter
T
2
T
1
T
2
I
5 (+)
5 (+)
6 (-)
6 (-)
170
Ω
Figure 5-15 Active and passive signals
Summary of Contents for SITRANS F M Intermag 2
Page 2: ......
Page 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Page 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...