Operation
5-15
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
Menu
code
Device function,
Parameters
Description
Factory setting
(display only)
Setting options
3.4
Total. net
Totalizer forw./rev.
3.4.1
Set net
Reset totalizer to zero and
stop or start
Cancel
1,2
Totalize
3
Cancel
1,2
,
Reset+Stop,
Reset+Start
1,2
,
Totalize
3
3.4.2
Engr. Unit
Engineering unit of volume
or
Engineering unit of mass
When you select a unit
of mass, entry of the
actual medium density is
mandatory (menu
3.1.4
).
DN 2...12: l
DN >12:
m
3
l, m
3
, ft
3
, Gal,
ImpGal
kg, t, lb
3.4.3
Lo Alarm Limit
Lower alarm limit
(the lower alarm limit must
be smaller than the upper
limit.)
-1 000 000 l
-10
8
to +10
8
3.4.4
Hi Alarm Limit
Upper alarm limit
(the upper alarm limit must
be greater than the lower
limit.)
+1 000 000 l
-10
8
to +10
8
3.4.5
Hysteresis
Hysteresis for limit values
0 l
0 to +10
8
1
Parameters apply only to devices with 20 mA/Hart communication and digital output 2
(7ME5033-0xxxx)
2
Parameters apply only to devices with 20 mA/Hart communication and digital input
(7ME5033-2xxxx)
3
Parameters apply only to devices with PROFIBUS communication
(7ME5033-1xxxx)
Summary of Contents for SITRANS F M Intermag 2
Page 2: ......
Page 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Page 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...