Operation
5-50
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
The DIAGNOSIS part contains diagnostic messages of the device
and has the following structure (Bit 0 = LSB, Bit 7 = MSB):
Bit No.
Description
Meaning
Cause (local message)
0
DIA_HW_ELETR
Hardware failure of the
electronic
”Com.--Module failure”
Communications module cannot be addressed
electronic
Sensor failure
(sensor failure)
”Measurement module failure”
4
DIA_MEM_CHKSUM
Memory error
”Memory failure”
5
DIA_MEASUREMENT
Measurement failure
”Flow measurement failure”
_
”Tube empty”
”Measuring range overflow”
Table 5-14
Format of the diagnostic data, Byte 5 (=DIAGNOSIS Byte 1)
Bit No.
Description
Meaning
Cause (local message)
0 ... 6
reserved
---
7
EXTENSION_
AVAILABLE
More diagnosis
information is available
More diagnosis information available (here:
Local messages see DIAGNOSIS--Byte 1)
Table 5-15
Format of diagnostic data, Byte 8 (=DIAGNOSIS Byte 4)
The bits in DIAGNOSIS usually take over the function of a group
message which can then be split up in detail using the local
messages.
The EXTENSION_AVAILABLE bits indicates that other diagnostic
information is available, basically details of the DIAGNOSIS
message. This information is displayed locally.
The measured value-related messages are transmitted additionally
cyclically with the measured values in the status byte (Substatus).
Examples for telegrams with diagnostic data (Ext_Diag_Data):
All diagnostic bits are reset in the initial state. When the ”Tube
empty” event occurs (byte 5, bit 5 set), this gives the following
diagnostic data:
Diagnosis Byte No.
1
2
3
4
5
6
7
8
Value
(hex)
08
FE
01
01
20
00
00
80
Meaning
s. Table 5-12
Header
Diagnosis
Summary of Contents for SITRANS F M Intermag 2
Page 2: ......
Page 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Page 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...