Operation
5-31
SITRANS FM Intermag 2 Flow Transmitter
Operating Manual - A5E00102775-03
Total. forwards
You have the following setting options in this menu:
menu item
3.2
S
Set
, menu item
3.2.1
In this menu you can reset the forwards totalizer to zero and restart
if necessary.
If the totalizer overruns the optical display of the totalizer stops at
“999999999” but the device continues counting internally. You have
to set a greater unit to be able to read the current totalizer reading
again.
You have the following individual setting options:
-
Reset+Stop: the totalizer is set to 0 and held.
-
Reset+Start: the totalizer is set to 0 and restarted
1,2
-
Totalize:
the totalizer is started
3
S
Unit
, menu item
3.2.2
With this function you can select physical units or switch from one
unit to another. In the latter case the device converts the previous
totalizer reading into the new unit.
S
Alarm limit
, menu item
3.2.3
Here you can set all values from 0 to +10
8
. You can assign reaching
or exceeding this limit value as a signal to a digital output.
Total. reverse
You have the following setting options in this menu:
menu item
3.3
S
Set
, menu item
3.3.1
In this menu you can reset the reverse totalizer to zero and
restart if necessary.
If the totalizer overruns the optical display of the totalizer stops at
“999999999” but the device continues counting internally. You have
to set a greater unit to be able to read the current totalizer reading
again.
You have the following individual setting options:
-
Reset+Stop: the totalizer is set to 0 and held.
-
Reset+Start: the totalizer is set to 0 and restarted
1,2
-
Totalize:
the totalizer is started
3
S
Unit
, menu item
3.3.2
With this function you can select physical units or switch from one
unit to another. In the latter case the device converts the previous
totalizer reading into the new unit.
S
Alarm limit
, menu item
3.3.3
Here you can set all values from 0 to -10
8
. You can assign reaching
or exceeding this limit value as a signal to a digital output.
1
Parameters apply only to devices with 20 mA/Hart communication and digital output 2
(7ME5033-0xxxx)
2
Parameters apply only to devices with 20 mA/Hart communication and digital input
(7ME5033-2xxxx)
3
Parameters apply only to devices with PROFIBUS communication
(7ME5033-1xxxx)
Summary of Contents for SITRANS F M Intermag 2
Page 2: ......
Page 8: ...SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 I 4 ...
Page 111: ...Appendix 7 3 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 113: ...Appendix 7 5 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 114: ...Appendix 7 6 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 115: ...Appendix 7 7 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 116: ...Appendix 7 8 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 117: ...Appendix 7 9 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 118: ...Appendix 7 10 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 119: ...Appendix 7 11 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 120: ...Appendix 7 12 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 128: ...Appendix 7 20 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...
Page 129: ...Appendix 7 21 SITRANS FM Intermag 2 Flow Transmitter Operating Manual A5E00102775 03 ...