E382 E407E 061 01 01 02
Siemens-Elema AB
Servo
…
Preventive maintenance
7 - 9
7
Lithium batteries
• Check manufacturing date for the lithium batteries
mounted onto PC 1771 and PC 1772.
The batteries must be replaced after 5 years.
Replacement is described in chapter ”Service
procedures”.
Safety inspection
• Make a visual inspection of the Servo
…
System for
external defects or damages. Replace defective or
damaged parts.
• Check the mains power cable and control cable and
their connections for damage.
• Perform a leakage current test. The leakage current
test is a standard procedure regulated by IEC 60 601-
1 or corresponding national standards. Allowable
values and test methods are defined in the standard.
The use of a leakage tester, e. g. Bender Safety
Tester 601/751 or equivalent is recommended.
• Check that a ”Servo
…
Ventilator System – User's
manual” and a ”Servo
…
Ventilator System – Brief
instructions” corresponding to the installed software
version is present. Also check that operating manuals
for all optional equipment connected to the Servo
…
System are present.
Completing the Preventive maintenance
• Perform a ”Pre-use check”. Refer to the
”Servo
…
Ventilator System – User's manual”.
• Perform ”Function checks” on the optional
equipments connected to the Servo
…
System.
Refer to the operating manuals for these optional
equipments.
• Note in the Servo
…
System software (Biomed) and
also in a Servo
…
System log book that a Preventive
maintenance has been performed.
Summary of Contents for Servo i
Page 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Page 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Page 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Page 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Page 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Page 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Page 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...