Siemens-Elema AB
Troubleshooting
Servo
…
E382 E407E 061 01 01 02
6 - 10
6
Breathing
10001 BATTERY_MIN_VOLTAGE
1.
Replace battery on PC1771 Control.
10002 BRE_I2C_ERROR
1.
Replace PC1771 Control.
10003 BRE_FATAL_MEMORY_ERROR
1.
Restart the Servo
…
.
2.
Replace PC1771 Control.
Panel
20001 PANEL_DEVICE_ERROR
1.
Replace PC1777 Panel.
20002 PANEL_BACKLIGHT_BROKEN
1.
Replace backlight lamps
2.
Replace PC board Backlight Inverter
20003 PANEL_BUTTON_STUCK
1.
Check user interface membrane buttons.
2.
Replace the touch screen.
3.
Replace PC1777 Panel.
20004 PANEL_AUDIO_FAILED
1.
Replace the loudspeaker
2.
Replace PC1777 Panel.
Exp Flow Meter
40001 Technical error in Exp flow meter
(More details in the log)
1.
Replace the expiratory cassette
2.
Replace PC1784 Expiratory channel.