Preventive maintenance
Servo
…
E382 E407E 061 01 01 02
7 - 6
7
On each one of the two gas modules:
• Open the hatch and pull out the nozzle unit.
• Carefully put the new plastic nozzle unit in position
into the gas module and close the hatch.
• Discard the old nozzle unit.
Wait 10 minutes before connecting pressure to the
gas modules.
• Re-assemble all parts inside the inspiratory section.
Do not mount the inspiratory section cover.
SVX9005X
On each one of the two gas modules:
• Unscrew the two screws (1) and lift off the
cover (2).
• Remove the rubber sealing (3) from the old filter (4)
and discard the old filter.
• Mount the rubber sealing on the new filter and put
the new filter into the cover (2).
• Mount the cover on the gas module.
All disposable parts must be discarded according to
local regulations and in an environmentally safe way.
SVX9124X
@
=
I
=
B
Note
: When assembling, make sure that the connector
muff seals around the inspiratory pipe (1).
The connector muff must not be pushed too far onto
the nozzle units (2).
SVX9145X
=
I
I
Summary of Contents for Servo i
Page 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Page 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Page 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Page 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Page 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Page 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Page 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...