Description of functions
3 - 3
3
Patient unit
Inspiratory section
Functional Main Blocks diagram marking: “
I
”
The main block I
NSPIRATORY
S
ECTION
conveys the
breathing gas from its gas inlets for Air and O
2
supply
to the patient breathing system. It comprises the
following main functions:
• G
AS
M
ODULES
– A
IR
AND
O
2
.
• C
ONNECTOR
M
UFF
.
• I
NSPIRATORY
P
IPE
.
• O
2
C
ELL
.
• T
EMPERATURE
S
ENSOR
.
• I
NSPIRATORY
P
RESSURE
T
UBE
.
• S
AFETY
V
ALVE
.
• I
NSPIRATORY
O
UTLET
.
• PC 1780 P
NEUMATIC
B
ACK
-P
LANE
.
Gas modules – Air and O
2
The Air and O
2
G
AS
M
ODULES
regulates the inspiratory
gas flow and gas mixture.
1. Filter
2. Inspiratory valve temperature sensor
3. Supply pressure transducer
4. Flow transducer (Delta pressure transducer and net)
5. Nozzle unit with valve diaphragm
6. Inspiratory solenoid
The G
AS
M
ODULES
are factory calibrated. Each G
AS
M
ODULE
must not
be disassembled further than
described in chapter ”Preventive maintenance”.
Gas inlet
The gas inlet nipples are quick-couplings where gas
supply is connected to the ventilator. The design of
the inlet nipples and their color markings vary
according to the standard chosen.
Gas is to be connected from hospital central gas
supply or from gas cylinders. The Air supply may be
connected from a compressor for medical air.
Filter
The F
ILTER
protects the ventilator from particles in the
gas delivered to the G
AS
M
ODULES
. The filter must be
replaced during the ”Preventive maintenance”.
The filter housing and the filter cover are provided
with matching guide pins. These guide pins prevent
mounting of the filter cover (with gas inlet nipple) on
the wrong module.
A non-return valve for the gas inlet is located in the
filter cover. This valve will suppress short pressure
drops in the gas supply.
The non-return valve is also designed to slowly
evacuate compressed gas from the module, if the gas
supply to the module is disconnected.
Inspiratory valve temperature sensor
The temperature of the supplied gas is measured by
the I
NSPIRATORY
V
ALVE
T
EMPERATURE
S
ENSOR
. This sensor
is situated in the gas flow.
The output signal from this sensor is used to
compensate for the gas density variations due to
temperature.
Supply pressure transducer
The pressure of the supplied gas is measured by the
S
UPPLY
P
RESSURE
T
RANSDUCER
.
The output signal from this transducer is amplified. It
is then used to calculate the absolute pressure of the
gas to compensate for gas density variations due to
pressure.
Flow transducer
The gas flows through a net (resistance) which
causes a pressure drop. The pressure is measured on
both sides of this net and the differential pressure
value is then amplified.
4
3
5
6
2
1
SVX9003X
Summary of Contents for Servo i
Page 1: ...Servo Service Manual E382 E407E 061 01 01 02 ...
Page 2: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 2 Notes ...
Page 8: ...Siemens Elema AB Important Servo E382 E407E 061 01 01 02 1 8 Notes 1 ...
Page 16: ...Siemens Elema AB Introduction Servo E382 E407E 061 01 01 02 2 8 2 Notes ...
Page 44: ...Siemens Elema AB Service procedures Servo E382 E407E 061 01 01 02 5 10 5 Notes ...
Page 64: ...Siemens ElemaAB Preventivemaintenance Servo E382 E407E 061 01 01 02 7 10 7 Notes ...
Page 65: ...E382 E407E 061 01 01 02 Siemens Elema AB Servo Index 8 1 8 8 Index Alphabetic index 8 2 ...