To separate the front panel section from the rear
cover:
• Disconnect the control cable (1).
• Remove the screws (2).
• Lift off the rear cover from the front panel section.
• All parts inside the front panel section are now
accessible.
Note
When replacing PC 1777 Panel, it can be necessary to
re-install the system software. For further information
refer to section ”Replacing PC boards” in this chapter.
SVX9100X
With power supply connected to the Servo
…
System, there are energized electrical components
inside the unit, e. g. the backlight lamps that are
supplied with 660 V by the Backlight Inverter.
All personnel must exercise extreme caution if
fault tracing or adjustments are performed with
power supply connected and with the user
interface rear cover removed.