Siemens-Elema AB
Preventive maintenance
Servo
…
E382 E407E 061 01 01 02
7 - 4
7
• Lower the locking catch (1) and lift off the O
2
cell.
• Disconnect and lift off the inspiratory pressure
tube (2).
• Press the two latches (3), one at each side of the
safety valve housing, and lift off the inspiratory pipe
(4) including the connector muff (5).
• Replace the filter (6) for the inspiratory pressure
transducer. Make sure that the filter is correctly
seated into the rubber ring.
Note
: This filter may already be replaced as a part of
the ”Extended cleaning of Inspiratory channel”
performed prior to the Preventive maintenance.
SVX9138X
=
I
SVX9139X
@
B
C
SVX9140X
D