Siemens-Elema AB
Disassembling and assembling
Servo
…
E382 E407E 061 01 01 02
4 - 6
4
To remove the expiratory cassette compartment (1):
• Disconnect the connectors (2 and 3) from PC 1784.
• Disconnect the tubes (4 and 5) from PC 1784.
• Disconnect the inspiratory pressure tube from the
inspiratory pipe inside the inspiratory section (not
visible in this illustration).
• Remove the two screws (6).
• Lift off the cassette compartment (1).
All parts behind the cassette compartment are now
accessible, e. g. the:
• PC 1780 Pneumatic back-plane (7).
• Safety valve pull magnet (8).