Revision Date: 3/13/2019
4
7.
Press the blue “system off” button on the Siemens scanner control box located on the
wall next to the MR Scanner window
How to Turn on the MRI Scanner/Computer
1.
You must first press the “system on” button on the Siemens scanner control box located
on the wall next to the MR Scanner window
2.
It will take approximately 15 minutes for the system to completely boot up
3.
After a successful reboot you will hear 3 beeps
MR Scanner Error Save Log
1.
If you ever have a scanner problem that requires you to reboot, it is very important to
make a MR Save Log entry BEFORE you reboot the system
2.
This log entry helps the Siemens engineer isolate the scanner problem and expedite the
system repair. To make a MR Save Log entry, you must do the following:
At the top of the screen – click system, then control, then tools
Click on “Save system log files”
o
You will see a box pop up with script on a black background, LEAVE
THIS ALONE!
o
Another box will pop up where you can input the problem
o
Fill out the “User” section and then click OK
o
You must wait until the black pop box with script closes before you can
proceed to reboot, scan etc
3.
Please email [email protected] with details of the problem