Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 7 of 48
1.10.2 Electrical :
Connect oscilloscope at D915.TP.kV.
Set the exposure parameters as 60kV 16mAs.
Release an exposure.
The recorded waveform will read 2.9V - 3.1V.
Repeat with 100kV 10mAs.
The recorded waveform will read 5V +/-0.25V.
Lower kV indicates code in kVsoll / kV sensing / kV-regulator OR high mA.
Higher kV indicates code in kVsoll / kV sensing / kV regulator.
1.11 mA/mAs Measurement :
mAs measurement could be carried out either by mAs-meter or by electrical
measurement.
1.11.1 mAs-meter :
Remove mAs link on D800 PCB on the Single tank.
Insert mAs-meter leads in the banana sockets mA+ and mA- on D800.
Set meter at, 200mAs and in the mAs mode.
Set exposure parameters as 60kV 32mAs.
Release an exposure.
The meter will read 31 – 33 mAs.
Repeat for exposure parameters 100kV 40mAs.
The meter will read 39 to 41mAs.
1.11.2 Electrical :
Connect oscilloscope at D915.TP.JR.
Set exposure parameters 60kV 16mAs.
Release an exposure.
The recorded waveform will read 3 – 3. 2V.
The exposure time will be 95 - 105mS.
Lower mA indicates code in mAsoll / mA sensing / mA regulator circuit.
Higher mA indicates code in mAsoll / mA sensing / mA regulator or Low kV.
Incorrect mAs indicate code in timer circuitry.
1.12 Calibration
:
It is advisable to re-calibrate the unit in case of undershoot or overshoot observed in
the mA wave shape.
Recalibration is required when the single tank is replaced.