4.5.3 Code 05 : kVist <> 0
START
KVIST = 0 ?
(NO SPIKES)
MAKE THEIR
PROPER
CONNECTION
X8 & X1
(ON D915 & D800) ARE
CONNECTED
PROPERLY ?
SWITCH OFF THE UNIT
R=20 Kohms AT
X8.1,X8.2 & X8.3
REPLACE D800 PCB
ON THE SINGLE
TANK.
Yes
REPLACE D915
STOP
No
Yes
No
YES
NO