Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 42 of 48
4.6.7 Code 18 : Premature termination of Exposure
START
PRESS KV mAs KEY.
ERROR IS REMOVED.
PRESS AND HOLD EXPOSURE RELEASE
SWITCH TILL EXPOSURE GETS
TERMINATED.
STOP