4.6.12 Code 01 : Err signal on D950A active
Start
Is tp test point on D950A
low?
Replace D950A.
NO
Is Vc>3.35v (A test point on
D950A?)
Yes
Yes
NO
Whether IGBT V3 is
Short circuited.
Stop