Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 31 of 48
4.5 Standby
Codes
4.5.1 Code 02 : +15 V Supply Code
START
REMOVE CONNECOR X15 ON D915
& SWITCH ON THE UNIT
M15 V ON X15
CONNECTOR
VOLTAGE=
13.5 TO 16.5 V ?
YES
REPLACE D915 PCB
NO
ADJUST BY PRESET ON
D980. ELSE
REPLACE D980 PCB
STOP