4.6.6 Code 17 : Exposure terminated by Backup Timer
START
MONITOR JR ON D915 &
mAS ON D800 PCBS
ARE PULSES ?
MULTIPLY JR & PULSE
DURATION OF J1.9
IS
IT EQUAL TO mAS
?
ADJUST FREQUENCY AT J1.9 BY
VARYING P1
TAKE AN EXPOSURE AND
NOTE THE TIME
TIME EXCEEDS
EXPOSURE TIME ?
REPLACE D915 PCB
STOP
YES
Yes
NO
ERROR CONTINUES
?
NO
Yes
NO
Yes
No