Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 34 of 48
4.5.4 Code 06 : JR <> 0
START
MONITOR JR & SWITCH ON
THE UNIT
IS
JR = 0 & NO
SPIKES ?
REPLACE D915
ENSURE ALL CONNECTORS ARE
CONNECTED.
YES
STOP
R=20 OHM AT X8.5 &
X8.6
REPLACE D800.
IF THE ERROR CONTINUES,
REPLACE SINGLE TANK AND
CALIBRATE THE UNIT.
CHECK X8
CONNECTOR ON D915.
No
SWITCH OFF THE UNIT
No
Yes
Yes
REPLACE D915
STOP