Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 38 of 48
4.6.3 Code 13 : Iheiz > Imax OR JR > Jrmax
START
MONITOR IH AND JR ON
D915 PCB
IS
IH MAX ERROR
?
DECREASE BY VARYING
POT P5 AND P6
ERROR CONTINUES ?
TAKE AN EXPOSURE
MONITOR I FOR
PULSES
REPLACE D801 PCB
REDUCE OVERSHOOTS
IN JR BY PRESET P5
ELSE REPLACE D915
NO
YES
REPLACE D915 PCB
YES
STOP
IS JR>JRmax ?
YES
NO
No
YES
NO