Maximum value of mAist above which Code is displayed is 180mA.
14
KVist < kVsoll
kVist is continuously polled during exposure. The value of kVist should be
greater than 85% of kVsoll
15
JR < JRS
JR is continuously polled during exposure. The value of JR should be greater
than 50% of JRS.
17
Backup Timer
If regular means of terminating exposure fails and exposure gets terminated
by the backup timer.
18
Premature Exposure
Termination
Exposure Release Switch is released before the exposure is terminated by
the mAs Integrator.
21
I > Iheiz
Same as Code 04, but during Exposure.
22
Maximum Preparation Time
This Code is displayed if First step is pressed for more than 15 sec.