4.6.8 Code 21 : Iheiz > Iheiz maximum
START
MONITOR IH ON D915 PCB
DECREASE BY VARYING
POT P5 AND P6
ERROR CONTINUES ?
TAKE AN EXPOSURE
IS THERE ANY SPIKE ?
NO
REPLACE D915 PCB
YES
STOP
No
REPLACE D801
YES