Service Manual
Service Instructions
Siemens Ltd. Med Division India
Version 4.0
Page 25 of 48
4 Specific Code Handling
Note :Please Ensure that the Calibration of the unit is done properly before
starting with the code Handling.
4.1
Unit not switching ON
Press the On Switch
on the
Top Panel
START
MAKE MAINS
SWITCH ON
PRESS SWITCH
I
ON
THE TOP PANEL
UNIT ON ?
IF
(ON/OFF SWITCH IS
Found faulty)
THEN REPLACE D936
ELSE
REPLACE D801
yes
No
Yes
No
LED V50 ON on D801
Yes
No
CHECK MAINS SUPPLY
Yes
B
Relay K5 and then K4 on
D801 ON
No
Yes
Contactors NS,CS
operates?
Check 220V on
D801X50.1-X50.3
Ensure relays K4,CS1
are operating.
No
Yes
B
Are Transformer O/P?
Check Voltages on D801
between terminals X1.6-X1.7
should be 24V~.
Replace Transformer
A
No
IF (ON/OFF SWITCH IS
found faulty) THEN
REPLACE D936 ELSE
REPLACE D801.