27
Siemens AG Healthcare Sector
11/12
Technical data
Line power supply
For all products, line power has to be supplied via an on-site system contact or via another
multipole shut-down mechanism. Room installation has to be in compliance with
VDE 0100-710
1
.
1. In all countries, compliance with local and national legal regulations is required.
However, we strongly recommend compliance with the regulations described herein – to the extent per-
mitted by relevant local and national laws – in order to ensure the safety of operating personnel, patients,
and third parties.
Power rating plate
XQ gradients
Values
Tolerance
Voltage
380 V, 400 V, 420 V, 440 V,
460 V, 480 V
±10%
Frequency
50/60 Hz
±1 Hz
Connection value
110 kVA
On-site protection
160 A
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...