Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
16
Gradient stray field distribution relevant for assessing exposure to
MR workers
The diagram shows the magnetic stray field distribution of the gradient system along the
patient axis of the MAGNETOM Skyra with the XQ gradient system according to the
requirements of the IEC 60601-2-33 standard.
The magnetic fields generated by each of the three orthogonal gradient axis are calculated
on a coordinate grid on a virtual cylinder surface. The cylinder encompasses the patient
axis and starts in the magnet isocenter.
The grid is defined by:
•
a point pattern on the surface of a virtual cylinder with a diameter of 70 cm (i.e. max.
bore liner);
•
points on circles on the cylinder surface perpendicular to the cylinder axis with an
azimutal step width of 10 degrees (i.e. 36 points at the circumference) and an axial step
width of 4 cm.
The field values are calculated at max. currents that can be handled by the gradient ampli-
fiers. At each space point the field vectors generated by each gradient are calculated,
superimposed and then their magnitudes are derived. The max. magnitude field value that
can be found on each circle is selected and plotted along the z-axis.
By dividing the maximum field values by the shortest rise time, the dB/dt values can be
derived.
Shortest rise time
Max dB
/
dt
XQ gradient
225 µs
60 mT/225 µs = 267 T/s
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...