Operator Responsibilities for Maintenance
7
© Siemens, 2009
M7-000.664.01.04.02
Page 7 of 14
MAGNETOM
03.12
CS SD MR
Electrical tests
Protective conductor
MRSC (MRWP)
Protecting personnel from elec-
tric shocks. (The protective con-
ductor resistance can change
during system operation, i.e.
through oxidation, corrosion,
and cable breaks.)
Protective conductor resistance
of the MRSC (MRWP)
Annually
Protective conductor sys-
tem
Protecting patients and person-
nel from electrical shock. (The
protective conductor resistance
can change during system oper-
ation, i.e. through oxidation,
corrosion, and cable breaks.)
Protective conductor resistance
of the entire system
Every
2 years
Function test
Emergency power off cir-
cuit
2
Safety of patient, personnel, and
system
Checking function of the power
circuit
Annually
Fix patient table (option) Safety of patient, personnel, and
system
Inspect table movement and
end switches
Operation of the emergency
stop and safety switches
Operation of the emergency
release
Checking tolerance limits
Marking of the hazard area
Annually
Object or function
Reason
What is checked:
Interval
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...