Safety information for the system owner
0.0
A.1-7
0.0
syngo MR D13
0.0
A.1
Quench emergency plan
A.1
It is strongly recommended that the system owner establishes
an emergency plan in case gaseous helium escapes into the
examination room.
A.1
The emergency plan should include the following information:
A.1
■
Rescue scenarios that can be practiced with personnel
■
Room-related conditions
■
Rescue personnel (safety personnel, paramedics and fire-
men)
Quench
A.1
During a quench, the super-conductivity of the magnet is sus-
pended. The energy of the magnetic field is converted into heat.
The magnet field strength falls off to 20 mT within approx.
20 seconds. The liquid helium (coolant) boils off during this pro-
cess and is released to the outside via the exhaust vent line. The
escape of gaseous helium via the exhaust line is rather noisy
(hissing, gurgling).
A.1
A quench may occur as follows:
A.1
■
Start-up of the MR system (ramping up or filling the mag-
net)
■
An accident (earthquake, fire, etc.)
■
Spontaneously without any obvious external reason (highly
unusual)
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...