Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
18
Spatial distribution of the gradient field at different positions
inside the gradient coil
The diagram shows the spatial distribution of the maximum magnitude values of the vec-
tor sum of the field components generated by each of the three gradient units simulta-
neously at positions on virtual cylinders coaxial with the patient axis with diameters of
0.2 m, 0.4 m, and 0.6 m.
The virtual cylinders have the same length as the gradient coil. In the cylinder axis direc-
tion the points have a separation of 0.02 m.
The area shaded in grey indicates the length of the gradient coil.
z [m]
|B| [T]
0.20 m
0.40 m
0.60 m
0.0
0.5
1.0
1.5
2.0
0.00
0.01
0.02
0.03
0.04
Magnet isocenter
End of gradient coil
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...