0.0
Safety instructions
A.1-8
0.0
MAGNETOM Family - Safety Information
0.0
Defective vent line
A.1
If the vent line fails in part or fully, gaseous helium will enter the
examination room. In this case, the air conditioning unit will not
be able to ensure sufficient air exchange and the following haz-
ardous situations may arise:
A.1
■
Poor visibility due to strong fog in the upper area of the
room
■
Rise in pressure in the examination room
■
Hypothermia and risk of suffocation (e.g. in case of large
leaks)
A.1
Due to such hazardous conditions as acute hypothermia and
suffocation, rescue attempts must be performed by at least
two persons.
Persons not directly involved in the rescue should leave the
examination room as well as adjacent rooms.
A filter (gas mask) without its own oxygen supply does not
protect against suffocation by helium.
An emergency plan must be established to ensure correct
conduct under such hazardous conditions.
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...