21
Siemens AG Healthcare Sector
11/12
MR compatibility data sheet
RF data
Peak power of transmitter amplifier
Channel 0
Channel 1
40.0 kW
≥
16.0 KW
≥
24.0 KW
Transmitter bandwidth
800 kHz
Receiver bandwidth
500 Hz–1 MHz (for each channel)
RF transmit coils
Body coil
Max. applied RF field B1+
24 µT
Max. specified B1+ rms
3.6 µT
Distance to isocenter, where RF transmit field is reduced by
3 dB
0.15 m
10 dB
0.27 m
CP Extremity coil
Max. applied RF field B1+
47 µT
Max. specified B1+ rms
9 µT
Tx/Rx 15-Channel Knee
coil
Max. applied RF field B1+
47 µT
Max. specified B1+ rms
8 µT
Tx/Rx Knee 15 MR coil
Max. applied RF field B1+
47 µT
Max. specified B1+ rms
8 µT
Tx/Rx CP Head coil
Max. applied RF field B1+
41.1 µT
Max. specified B1+ rms
6 µT
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...