Safety information for the system owner
0.0
A.1-3
0.0
syngo MR D13
0.0
A.1
Electromagnetic fields
A.1
The 0.5 mT line in the examination room defines the controlled
access area of the RF field. For controlling access to this area
adequate rules must be established. Therefore, regard the
potential risks from the attraction of magnetizable objects or
from torque on such materials. Also consider persons inadvert-
ently entering the area who may be affected by the possible dys-
function of their medical implants such as pacemakers.
A.1
Outside the controlled access area of the RF field, electromag-
netic interferences meet the requirements according to
IEC 60601-1-2.
A.1
Regarding the static magnetic field, your MR system is continu-
ously operated in the normal mode (means not more than
3.0 T).
For China only
: If the static magnetic field is higher than
2.0 T, the system is operated in the first level mode.
A.1
The limits of exposure of MR workers to static and time-varying
magnetic fields may be regulated by local laws. Special precau-
tions are necessary for pregnant MR workers, although no epi-
demiological evidence of any negative health effects currently
exists (local laws may apply). It might be that the limits do not
apply for pregnant MR workers. Furthermore, it may be required
in some countries that the “member of the public” limit is
applied to the fetus.
A.1
Noise development
A.1
The exposure of MR workers to noise may be regulated by local
laws. The operator must ensure that the sound level at the oper-
ating console is limited in compliance with local rules for the
safety of the physician and the MR worker.
A.1
For MR equipment that produces noise louder than 99 dB(A),
the sound pressure level is measured according to NEMA MS 4.
A.1
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...