Operator Responsibilities for Maintenance
Quench system
Safety of patient, personnel, and
system
Visual inspection of the quench
tube
•
Check for water in the quench
valve
•
Check quench tube and
isolation
•
Check quench tube outlet for
obstructions
•
Check outlet rain cover
•
Check protective mesh
•
Check quench tube outlet,
restricted area and warning
signs
Annually
Operating manuals
Safety of patient, personnel, and
system
Required operating manuals are
available and legible
Annually
User icons, button label-
ing, warning labels
Safety of patient, personnel, and
system
Visual inspection of the user
icons, button labeling, and
warning labels
Annually
Exclusion zone for mag-
netic field
Prevent hazards caused by mag-
netic field
The 0.5 mT zone is marked
The examination room is identi-
fied with the warning “strong
magnetic field”
Annually
Object or function
Reason
What is checked:
Interval
Summary of Contents for MAGNETOM Skyra
Page 2: ......
Page 4: ......
Page 8: ...Siemens AG Healthcare Sector 11 12 Introduction Page 4 of 4 ...
Page 12: ...0 0 Safety instructions A 2 0 0 MAGNETOM Family Safety Information0 0 ...
Page 37: ...A 1 ...
Page 39: ...MR MAGNETOM System Owner Manual MR compatibility data sheet Skyra ...
Page 62: ...Siemens AG Healthcare Sector 11 12 MR compatibility data sheet 24 ...
Page 63: ...MR MAGNETOM System Owner Manual Guidance and manufacturer s declaration EMC Skyra ...
Page 73: ...MR MAGNETOM System Owner Manual Technical data Skyra ...
Page 102: ...Siemens AG Healthcare Sector 11 12 Technical data 30 ...
Page 103: ...MR MAGNETOM System Owner Manual Location of labels Skyra ...
Page 105: ...3 Siemens AG Healthcare Sector 11 12 Location of labels Patent label ...
Page 107: ...5 Siemens AG Healthcare Sector 11 12 Location of labels Safety certificate label U S A only ...
Page 108: ...Siemens AG Healthcare Sector 11 12 Location of labels 6 ...
Page 123: ...MR MAGNETOM System Owner Manual Disposal Skyra ...