Commissioning
6.3 Empty Pipe Set
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
87
Using the MTYmatic command
You can repeat MTYmatic (performed during the Initial Makeup) to correct an inaccurate
Actual MTY setting if conditions do not allow you to repeat the Actual Empty procedure.
Note
IMPORTANT
Only use the MTYmatic procedure when the pipe is full.
To start MTYmatic:
1.
From [Channel Setup] scroll down to [Install Sensor].
2.
Press the <Right Arrow> to access the [Empty Pipe Set] option list.
6LHPHQV
&KDQQHO>@$%&
6HW(PSW\3LSH'HWHFWLRQ7KUHVKROG
,QVWDOO6HQVRU
6HQVRU0RGHO
+37
6HQVRU6L]H
'+
6HQVRU0RXQW0RGH
5HIOHFW
6SDFLQJ2IIVHW
1RUPDO
1XPEHU,QGH[
6SDFLQJ0HWKRG
6SDFHU%DU%1
/WQ9DOXHLQ
(PSW\3LSH6HW
!07<PDWLF
=HUR)ORZ$GMXVW
$FWXDO07<
,QVWDOO&RPSOHWH
6HW(PSW\
3.
Move the cursor next to [MTYmatic] press the <Right Arrow>.
4.
To invoke MTYmatic press <ENTER>.
Using the Set Empty command
Use [Set Empty] to enter a number that represents the signal strength level consistent with
an empty pipe. [Set Empty] uses non-linear scaling. There is no direct correlation between
the number you enter and any standard amplitude unit. If you set the number too low, the
meter may not detect a true empty pipe. If you set it too high, it could trigger the empty pipe
alarm, suspending flow measurement, even though the liquid is flowing.
To enter an Empty Pipe Alarm Threshold:
1.
From [Channel Setup] scroll down to [Install Sensor].
2.
Press the <Right Arrow> to access the [Empty Pipe Set] option list.
3.
Press <Up Arrow> to move the cursor to [Set Empty].
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...