
Appendix
A.4 Flowrate Calibration and Calibration Tables
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
215
To enter the Kc Factor
1.
To enable numeric entry press <Right Arrow>.
2.
Use the numeric keys to type the required Kc (as calculated above). Note that the Kc
value can be negative or positive. Enter the - or + sign first, then type in the calibrated
value.
3.
To store the data press <ENTER>. Note that Kc now appears in the right-hand column of
the [Calibrate Flow Rate] menu cell with its new value. Also note that this Kc value can be
viewed on the site printout.
Calibration Tables 1 through 3
SITRANS F 1010 instruments offer a unique methodology by which a particular flow
response of an instrument may be linearized or optimized by tabulating the results of a
series of calibration exercises or collected batch data points. Basically, the flow meter allows
the user to select any of a wide variety of system variables (flow rate, pressure, viscosity,
etc.) as a pointer into a table of calibration factors (up to 32). As the system variable is
updated, the value of the table’s output factors (or positive and negative flow) is re-evaluated
and used as a modifier for the current rate register. Note that the flow register is still zeroed
and Reynolds number compensated normally and these slope corrections are in addition to
these fundamentals.
Note
Kc is still active when this method is being used.
To install a Calibration table:
1.
The user selects a system variable that appears to correlate strongly with calibration
shifts observed.
2.
A table of values is formed comprised of the values that this index could assume over the
range of system operation. Remember, the tables created do not extrapolate beyond their
end points, they "clip."
3.
A calibration factor, a number usually close to 1.00, is entered as a positive and a
negative flow rate correction factor (termed PosFlow Corr and NegFlow Corr) for each of
the desired index points.
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...