Functions
7.13 Resistive Temperature Device (RTD) Calibration
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
137
7.13
Resistive Temperature Device (RTD) Calibration
The [RTD Calibrate] menu appears on all SITRANS 1010 models. Use this menu to calibrate
Temperature Sensors to an external standard. It is important to note that Siemens RTD
temperature sensors are factory-calibrated for high accuracy. We recommend that before
deciding to perform the calibration, check the current RTD reading in the [Diagnostics Data]
menu. You may find that you do not need to calibrate the sensor. In any case, make sure
that the temperature reading stabilizes before proceeding further. The [RTD Calibrate] menu
allows you to perform an external calibration, which can be accomplished either by data
entry of the current RTD temperature or by a 0°C (32°F) Ice-Bath procedure. You can switch
between the intrinsic and external calibration modes at any time.
Note
If you perform an external temperature calibration, you should mark and record the location
of each connector and sensor-cable. Once you have re-calibrated the temperature sensors,
changing the sensor/connector orientation established during the procedure may void the
calibration.
1.
From the [Meter Facilities] menu scroll to the [RTD Calibrate] menu.
2.
To access the [RTD Calibrate] menu press <Right Arrow>.
0HWHU)DFLOLWLHV
6LHPHQV
'XDO3DWK>@3DWK
&DOLEUDWH7HPSHUDWXUH6HQVRUV
3UHIHUUHG8QLWV
(QJOLVK
7DEOH6HWXSV
/RJJHU&RQWURO
0HPRU\&RQWURO
$QDORJ2XW7ULP
57'&DOLEUDWH
&ORFN6HW
566HWXS
1RQH>@
%DFNOLJKW
2Q
6\VWHP,QIR
Table 7- 11 RTD Calibrate Menu Structure
RTD 1→
Factory
User Cal
RTD Calibrate
RTD 2→
Factory
User Cal
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...