5.5 Sensor Installation
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
56
Operating Instructions, 01/2013, A5E02951520-AC
Selecting a location for the sensors
1.
Locate the sensors downstream from the center of the longest available straight run. A
location ten pipe diameters or greater downstream from the nearest bend will provide
adequate flow profile conditions.
2.
Do not, if possible, install the sensors downstream from a throttling valve, a mixing tank,
the discharge of a positive displacement pump or any other equipment that could possibly
aerate the liquid. The best location will be as free as possible from flow disturbances,
vibration, sources of heat, noise, or radiated energy.
3.
Avoid mounting the sensors on a section of pipe with any external scale. Remove all
scale, rust, loose paint, etc., from the location.
4.
Do not mount the sensors on a surface aberration (pipe seam, etc.)
5.
Do not mount sensors from different ultrasonic flow meters on the same pipe. Also, do
not run the sensor cables in common bundles with cables from other instrumentation.
You can run these cables through a common conduit ONLY if they originate at the same
flow meter.
6.
Never mount sensors under water, unless you order submersible units and you install
them in accordance with factory instructions.
7.
Avoid mounting sensors on the top or bottom of a horizontal pipe. The best placement on
a horizontal pipe is either the ten o’clock and two o’clock position for Reflect Mode, or one
sensor at nine o’clock and one sensor at three o’clock for Direct Mode. Mounting on a
vertical pipe is recommended only if flow is in the upward direction. When mounting on a
vertical pipe flowing in a downward direction make sure there is sufficient back pressure
in the system to maintain a full pipe.
R
①
Dual Path, Reflect Mount
②
Dual Path, Direct Mount
Figure 5-8
Sensor Alignment (Horizontal Plane)
Preparing the Pipe
1.
Pick a mounting location with the longest straight run. You must have easy access to at
least one side of your pipe. The pipe at the mounting location must remain full, even at
zero flow.
2.
Decide on your mounting mode (Direct or Reflect). Always use Reflect Mode whenever
possible. You may only need to use Direct Mode if your pipe is plastic.
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...