Connecting
5.5 Sensor Installation
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
64
Operating Instructions, 01/2013, A5E02951520-AC
2.
Slide sensor into a mounting frame back end first aligning the angled edge of the sensor
with the angled edge of the mounting frame. Keep sensor from making contact with the
pipe until it butts up against the mounting frame stop. Push sensor down to mate with
pipe.
3.
Tighten the sensor clamping screws to hold the sensor firmly in place.
Repeat procedure
for the other sensor.
4.
If installing a temperature sensor proceed to Mounting Temperature Sensor (Page 78). If
not, proceed to Sensor Wiring (Page 80).
5.5.4
Direct Mount
Sensor Installation using Mounting Frames, Spacer Bar and Spacing Guides
The combination of mounting frames, spacer bar and spacing guides is the recommended
way to mount Direct Mode sensors. The mounting frame establishes the axial alignment of
the sensors and allows you to remove and replace either sensor while preserving their exact
mounting location.
For Direct Mode mounting, a spacer bar is used to establish the distance between sensors
and a spacing guide to locate the sensors at the nine o’clock and three o’clock positions.
Should the distance between sensors be beyond the span of a spacer bar, a measuring tape
can be used. The Mylar spacing guide comes in various lengths and widths to accommodate
most pipe sizes.
Spacing Guide Sizes
Metric
English
5.08cm x 66.04cm
2" x 26"
5.08cm x 114.3cm
2" x 45"
10.16cm x 393.7cm
4" x 155"
15.2cm x 497.8cm
6" x 196"
Figure 5-15 Mylar Spacing Guide
1.
After receiving the spacing index from the Installation Menu, prepare the pipe surface
area where the sensors will be mounted.
2.
Degrease the surface and remove any grit, corrosion, rust, loose paint, etc.
3.
Make a note of the Number Index displayed in the [Install] menu. Check to ensure that
you have a matched set of sensors. They both should have the same S/N number but
marked with either an "A" or "B" (e.g., 100A and 100B).
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...