
Appendix
A.5 Setup Procedure for Wet-Flow Calibration
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
219
4.
Proceed to the [Install Sensor] menu and select the same sensor, mounting mode and
spacing offset that was selected for the water calibration.
5.
Re-space the sensors to the index position indicated by the flow meter.
6.
Scroll to the [Install Complete] menu cell and select [Transfer Install] from the option list.
Note
If [Transfer Install] does not appear in the option list then either the Estimated Vs or the
sensor size was improperly entered. In this case, recall the water calibration site and start
the procedure again at Step 1 above.
7.
For Multi-Path systems repeat Step 6 above for the remaining paths.
8.
The flow meter should now be operational at the new spacing location.
Note
Depending on the size of the pipeline, the change in the estimated sonic velocity (Vs) and
the repositioning of sensors, the flow meter may not operate out of Fault even if the spool
or pipe is filled with liquid. This can be expected when performing a Transfer Install for
liquefied gases or for clamp-on natural gas flow meters.
Saving New Transfer Install Site
1.
To save the Transfer Installed site, scroll to the [Channel Setup] menu and press the
<Right Arrow>. Press the <Right Arrow> again to select the [Save/Rename] menu cell.
2.
Use the numeric keys to rename the Transfer Installed site with the same site name used
above, but with a "T" appended to the end of the site name (e.g., 3CS40T).
3.
Press <ENTER> to store data.
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...