A.5 Setup Procedure for Wet-Flow Calibration
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
218
Operating Instructions, 01/2013, A5E02951520-AC
6.
The meter will perform a momentary "Makeup" routine that will take a few seconds and
then begin operation.
7.
Refer to your Operations Instruction manual for instructions on setting zero flow.
Note
Setting zero flow must be performed each time the sensors are installed. The zero
adjustment has no effect on the wet-flow calibration data or the calibration (Kc) factor.
8.
Using the arrow keys, scroll to the [Span/Set/Cal] menu location. Verify that the [Kc]
(calibration) factor matches the value indicated on the diagnostic data sheet.
9.
If you are measuring a liquid other than ambient water, select the [Liquid Class] menu cell
and <Down Arrow> from there to [Viscosity]. Enter the correct viscosity for the liquid you
will be monitoring.
10.
Setup is now complete. Press the <MENU> key twice to view the flow rate display. DO
NOT utilize the [Save Site] command when it appears.
11.
When measurements are completed, simply turn off the meter. DO NOT save the site.
This might contaminate the wet-flow calibration data already stored.
Note
Contact Siemens Technical Services Department
http://www.siemens.com/automation/service&support
) if any flow calibration data is
accidentally removed or overwritten.
Transfer Install Function
All SITRANS F 1010 flow meter operating systems (version 3.00.20 and greater) include the
installation facility called "Transfer Install." This function permits the sensors to be
repositioned while maintaining all calibration parameters and operation established during
the water calibration. The Transfer Install function allows the sensors to be optimally
positioned for a different fluid, without the need for a new Initial Makeup procedure.
Note
Prior to performing Transfer Install make sure that the water calibration procedure was
performed and a saved active site exists.
Transfer Install Procedure
To initiate the Transfer Install function, proceed as follows:
1.
In the [Application Data] menu press the <Right Arrow> to select the [Liquid Class] menu
cell. Scroll and highlight [Estimated Vs M/S] from the option list.
2.
Use the numeric keys to change the Estimated Vs to the Estimated Vs value of the
customer selected liquid.
3.
To enter new Estimated Vs value press <ENTER>.
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...