Troubleshooting
10.3 Test Facilities Graph Screen
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
Operating Instructions, 01/2013, A5E02951520-AC
165
[MinDamp #] Command
Pressing the <1> key will cause [MinDamp #] to appear on the command line at the lower
left-hand corner of the screen. The number listed to the right of the command code
represents the exponent in the meter exponential averaging routine, where the larger the
number the greater the digital averaging. Pressing the <+> key will increase the damping
value. Likewise, pressing the <-> key will decrease the damping value.
To exit this mode, press the <0> key on the keypad.
[MaxDamp #] Command
Pressing the <2> key will bring up the [MaxDamp #] command. The function of this
parameter is similar to the [MinDamp #] command described above; however, the two
parameters interact in the following manner. The MinDamp value must not exceed the
MaxDamp value; therefore increasing the MinDamp value above the previous MaxDamp
value will set both parameters to the same value. In most cases, it is preferred that both
damping parameters be set to the same value, however, in cases where rapid response to
changes in liquid sound velocity for flow rate is required, the two values may be set
differently. In this situation the meter will use the MaxDamp value when conditions are
stable, but then switch to a faster damping value (limited by MinDamp) when a significant
change in sound velocity or flow rate is perceived.
To exit this mode, press the <0> key on the keypad.
To access the Digital Damping Control using the Test Facilities Graph Screen, proceed as
follows:
Note
To use the Test Facilities Graph Screen you must have a working site.
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...