Appendix
A.4 Flowrate Calibration and Calibration Tables
FUS1010 IP65 NEMA 4X & IP66 NEMA 7
214
Operating Instructions, 01/2013, A5E02951520-AC
Some applications may require an output adjustment to match an official external reference.
The [Calibrate Flowrate] menu allows you to select a calibration mode. The right-hand
column shows the active calibration mode. You can select Intrinsic (factory) and Kc (Slope
Correction) Calibration. Selecting either of the external calibration modes will not eliminate
the Intrinsic (factory) calibration. You can use this menu cell to switch between Intrinsic and
Kc at any time.
Kc Calibration
For most applications, the measured flow range produces a linear meter response.
Therefore, the Kc (slope correction) calibration is the preferred method since it only requires
a single correction factor for all the flow rates encountered.
Note
Changing the calibration can cause profound changes in flow meter operating
characteristics. Use only the most respected flow standard to obtain a correction factor. The
percentage entered must provide an accurate and consistent shift across the entire flow
range anticipated for the application.
Kc Factor
To obtain the Kc factor, compare flow total data taken simultaneously from the flow meter
and a reference meter whose accuracy meets the required standard. Allow both meters to
accumulate flow total data long enough to average out any differences due to flow fluctuation
between the two meter locations. Compare outputs of the two totalizers to determine
percentage increase (+) or decrease (-) that is necessary to produce the best average
correlation between the flow meter and the reference standard.
Selecting the Kc Factor
When the [Kc] menu cell is selected, the flow meter imposes this percentage slope
adjustment of its rate output. Output data is zeroed and corrected for the Reynolds number
(flow profile compensated), however, a percent change in the rate output is imposed based
on the data entered in this cell. The number entered by the user is evaluated into a slope
correction factor by dividing it by 100 and algebraically adding it to 1. The resulting factor is
used as a multiplier on the rate register of the instrument. Thus an entry of -3% will multiply
the rate register by 0.97, for example.
To calculate Kc:
.F
$FWXDO5DWH
,QGLFDWHG5DWH
[
Summary of Contents for FUS1010
Page 224: ......
Page 225: ......
Page 226: ......
Page 227: ......
Page 228: ......
Page 229: ......
Page 230: ......
Page 231: ......
Page 232: ......
Page 233: ......
Page 234: ......
Page 235: ......
Page 236: ......
Page 237: ......
Page 238: ......
Page 239: ......
Page 240: ......
Page 241: ......
Page 242: ......
Page 243: ......
Page 244: ......
Page 245: ......
Page 246: ......
Page 247: ......
Page 248: ......
Page 249: ......
Page 250: ......
Page 251: ......
Page 252: ......
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......
Page 257: ......
Page 258: ......
Page 259: ......
Page 260: ......
Page 261: ......
Page 262: ......
Page 263: ......
Page 264: ......
Page 265: ...21614 C 1011NFPS 7 ...
Page 266: ...1011NFPS 7 C 21614 ...
Page 267: ...1011NFPS 7 C 21614 ...
Page 268: ...21614 C 1011NFPS 7 ...
Page 269: ......
Page 270: ......
Page 271: ......
Page 272: ......
Page 273: ......
Page 274: ...21614 C ...
Page 275: ......
Page 276: ......
Page 277: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 278: ...1012TN 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 279: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 280: ...1012TNH 7 INSTALLATION DRAWING 1010 SERIES TRANSDUCERS AND MOUNTING TRACKS 21614 C ...
Page 281: ......
Page 282: ......
Page 283: ......
Page 284: ......
Page 285: ......
Page 293: ...Glossary FUS1010 IP65 NEMA 4X IP66 NEMA 7 228 Operating Instructions 01 2013 A5E02951520 AC ...
Page 297: ...Index FUS1010 IP65 NEMA 4X IP66 NEMA 7 232 Operating Instructions 01 2013 A5E02951520 AC ...