Page EN-4
XCT-8 Sanitary Transducer– INSTRUCTION MANUAL
7ML19985HX61
English
Specifications
Measurement Range
• 0.6 m (2 ft) to 8.0 m (26 ft). Dependent on application conditions.
Temperature Range
• – 40 to 125 °C (–40 to 260 °F) at 0.6 m (2 ft)
Beam Angle
• 12°
Frequency
• 44 kHz
Environmental
• Location:
indoor / outdoor
• Ambient temperature:
– 40 to 125 °C (–40 to 260 °F)
• Altitude
2000 m maximum
• Pollution degree
4
Pressure
• 200 kPa (29 psi)
Temperature Sensor
• Internal
Enclosure
• PVDF with PTFE transducer face
• PTFE flange / ferrule
Mounting
• Base:
4” 3A compliant sanitary fitting clamp
• Conduit:
1” NPT thread or 1” BSP
Cable
• 2-wire twisted pair / braided and foil shielded, 0.55 mm
2
(20 AWG), PVC jacket
• Maximum cable run:
365 m (1200 ft)
Approvals
•
3A
•
CE
•
CSA
NRTL/C
Summary of Contents for Echomax XCT-8