UMSITRPDS3-1
Calibration And Maintenance
November 2010
7-15
MG
0
0353a
Differential Construction
Gauge Construction
1
2
3
4
Dimension A = 0.1" +/- 0.03 (2.6 mm +/- 0.75)
A
A
1
2
3
4
Enclosure rotation arrow (enclosure rotation reference point)
Ramp-shaped recessed area indicating enclosure rotation range
Enclosure setscrew, use 3/32 (2.5 mm) Allen screw
Tag plate
FIGURE 7-4 Measuring Cell Alignment and Insertion Depth
6.
Tighten the enclosure setscrew to 30.1 to 30.9 in lbs (3.4 to 3.6 Nm).
7.
Get the electronics module and connect the measuring cell board as described in Section 7.4.2.
8.
Install the electronics module as described in Section 7.4.2.
9.
Install the display as described in Section 7.4.1.
10.
Install the end cap by turning it clockwise until the O-ring contacts the enclosure. Turn the cap one
additional turn to compress the O-ring.
11.
Apply power to the transmitter; see Section 3 for wiring diagrams. Refer to the table in Section 7.4
and to Section 6 On-Line Configuration and Operation and configure the transmitter.
12.
Reinstall transmitter as described in Section 7.6 Transmitter Replacement.
Summary of Contents for 7MF4033 series
Page 48: ...Model 275 Universal Hart Communicator UMSITRPDS3 1 November 2010 2 26 ...
Page 86: ...Installation UMSITRPDS3 1 November 2010 4 34 ...
Page 152: ...Calibration and Maintenance UMSITRPSD3 1 November 2010 7 20 ...
Page 158: ...Circuit Description UMSITRPDS3 1 November 2010 8 6 ...
Page 197: ......
Page 198: ......
Page 200: ...Model Designations and Specifications UMSITRPDS3 1 November 2010 9 42 ...
Page 204: ...Glossary UMSITRPDS3 1 November 2010 10 4 ...
Page 208: ...Appendix A Online Configuration Map UMSITRPDS3 1 11 4 November 2010 ...
Page 216: ...Appendix C Elevation and Suppression Correction UMSITRPDS3 1 13 4 November 2010 ...