![SICK ZIRKOR Series Operating Instructions Manual Download Page 60](http://html.mh-extra.com/html/sick/zirkor-series/zirkor-series_operating-instructions-manual_1264209060.webp)
60
8024252/AE00/V1-0/2020-07| SICK
O P E R A T I N G I N S T R U C T I O N S | ZIRKOR200 Ex-D
Subject to change without notice
8
MAINTENANCE
8.6.4
Setting valves position
Fig. 34: Terminal cover of the Safe Area control unit
8.6.5
Setting the flow volume (19" rack)
On systems with 19" rack with integrated pneumatics, it is possible to set the reference or
test air volume on the electronics. A distinction is made here between the pump and
instrument air version:
●
Only the test air volume can be set for the pump version.
●
With an instrument air version, it is possible to set both the reference and test air
volume.
Fig. 35: 19“ rack - setting the flow volumes
Arrows show the position of the setting valves for reference air and test air
Setting valve
Reference air
30 l/h to 40 l/h
Setting valve
Test air
150 l/h to 180 l/h
Instrument air version
Pump version
.
A
Controller, test air
B
Controller, reference air
C
Controller, test air