Switching behavior
PNP switching to supply voltage V
S
Default settings in the device: no function, logic: not inverted (active
high)
Properties
•
Short-circuit pro temperature protected
•
Not electrically isolated from V
S
Electrical values
0 V ≤ V
out
1)
≤ V
S
(V
S
−1.5 V) ≤ V
out
≤ V
S
at I
out
2)
≤ 100 mA
1)
Output voltage.
2)
Output current.
NOTE
Assign the functions for the digital outputs in the device using a configuration tool, e.g.
the configuration software SOPAS ET.
13.4.10
Wiring digital outputs “External output 1” and “External output 2” of the device in the CDB650-204
Device = Lector621 = V2D621x-xxxxxYx (Y = B or C)
Load (e.g. PLC)
3
GND
V
out
For inductive load:
5
Device
1
Aux
(RS-232)
CDB650-204
CMC600
2
U
IN
*
“External
output
A
”
6
C
EXT. OUT
B
22
GND
6
Shield
4
Figure 47: Wiring external digital outputs “External output 1” and “External output 2” of the device in the CDB650-204 connec‐
tion module
1
Device
2
The optional CMC600 parameter cloning module is required in the connection module in order to be able to use the
additional external digital inputs and outputs of the device
3
Load (e.g. PLC)
4
Output voltage V
out
5
With inductive load: see note
6
Logical “External output” in the device
Inductive load
NOTE
Provide an arc-suppression switch at the digital output if inductive load is present.
b
Attach a freewheeling diode directly to the load for this purpose.
ANNEX
13
8022502/15NT/2020-02-11 | SICK
O P E R A T I N G I N S T R U C T I O N S | Lector621
87
Subject to change without notice