The device now ignores all further external read cycles. The device does not output
any read results via the host interface. The beeper confirms this process with an
ascending melody.
✓
The
Read Diagn
function is preselected as the first function (LED is lit).
2.
Press the arrow button (
◄
) repeatedly until the LED of the desired function lights
up. The device runs through all possible functions step-by-step without executing
them and then starts again from the beginning. The beeper confirms every step
with a tone.
3.
Press the return key (
p
) to confirm the selected function.
✓
The LED of the selected function now flashes and the beeper confirms start with a
double beep.
✓
The device executes the function and automatically returns to read mode after
5 minutes without button operation.
✓
The device automatically ends the
TeachIn
(for match code),
Auto-Setup
and
Autofo‐
cus
(Lector621 variant only) functions when the device has successfully read the
presented code. The assigned LED lights up green and the beeper confirms the
successful read with an ascending melody.
4.
To manually end (cancel) a function, press the return button (
p
) again and hold
for 3 seconds.
✓
The beeper confirms the change with an ascending melody. The
Ready
LED lights
up again.
✓
The device is ready for reading and is waiting for a read cycle.
NOTE
With the
Auto-Setup
function, make sure that a medium-high object is in the reading
field.
Table 5: Device functions (second display level for action with function buttons)
Display
LED
Color
Status
Read diagn
Lights up Blue
Test (reading diagnostics) selected
Flashing
Blue
Test started
TeachIn
Lights up Blue
Teach-in selected (default: Match code)
Flashing
Blue
Teach-in started
Lights up Green
Teach-in successful
Lights up Red
Teach-in unsuccessful (Default setting: match code, the
device was not able to teach in any code)
Auto-Setup
Lights up Blue
Auto-Setup selected
Flashing
Blue
Auto-Setup started
Lights up Green
Auto-Setup successfully quit
Lights up Yellow
Auto-Setup was partially successful
Lights up Red
Auto-Setup was unsuccessful
Autofocus
1)
Lights up Green
Function can be defined by user
Lights up Yellow
Function can be defined by user
Lights up Blue
Function can be defined by user
Lights up Red
Function can be defined by user