8012426/YWL2/3-0/2016-08| SICK
O P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER SP100
Subject to change without notice
PRODUCT DESCRIPTION
2
2.3.2
Voltage and purge air supply
Internal duct
pressure
[hPa ]
Connection and supply components
Purge air
Voltage
-50 ... +10
MCU-P + purge air hose DN25
-50... +30
Optional external purge air unit + purge air reduction
MCU-N
-50 ... +100
Adapter for instrument air
[1]
[1]
Instrument air onsite (free of dust, oil, moisture, non-corrosive)
-800 ... +2000
Non-return valve
[2]
for instrument air connection
1)
[2]
Part of sender/receiver unit for internal duct pressure up to +200 kPa
NOTE:
For temperatures above 220°C:
▸
Always use the optional external purge air supply and sender/receiver unit
DHSP-T4xxxNNXX type up to 400°C.
We recommend using the optional external purge air unit when the sender/receiver unit
is more than 10 m away from the MCU control unit.