Operating instructions Distance measuring device DL100 Pro – RS-422
Operation via Ethernet (Ethernet interface)
64
© SICK AG • Subject to change without notice • 8014753/ZVZ7/2018-03-22
Field "MF1, Threshold
distance underflow"
(continued)
Parameter
Description
Hysteresis
distance
Enter hysteresis for switching threshold for the multifunction
output MF1.
Read/Write access
• Read and write
Input value
• 1 ... 300000
Unit
• mm
Factory setting
• 10
Table 36:
Page "Parameter settings" – field "MF1, Threshold distance underflow"
Field "MF1, Threshold velocity
exceeded"
Requirements for the display
• Parameter "MF1 activation": Option "On"
• Parameter "Function": Option "Velocity"
Parameter
Description
Threshold
velocity
Enter switching threshold for the multifunction output MF1.
Read/Write access
• Read and write
Input value
• 0 ... 15000
Unit
• mm/s
Factory setting
• 5000
Distance
change
Choose the travel direction to be monitored.
Read/Write access
• Read and write
Input value
• 0: Increasing (positive values) / 1: Decreasing (negative
values) / 2: increasing and decreasing
Factory setting
• Increasing and decreasing
Table 37:
Page "Parameter settings" – field "MF1, Threshold velocity exceeded"