Teach function
Time [ms]
Switching on teach confirmation
2400
Signal level warning Q
1
: Teach Q
2500
Signal level warning Q
1
: Teach Q not
2600
Signal level warning Q
2
: Teach Q
2700
Signal level warning Q
2
: Teach Q not
2800
Teaching in the zero point
2900
Laser off
> 5,000
1
Centering limits. The near and far points that have been moved via centering must always lie within the
value range limits. Moved points must always be evaluated by the user.
8.4.3.2
In1 function
Use the In1 function for the following tasks:
•
Configure various device parameters
•
Switch the laser on and off at defined times
•
Activate device functions
•
Activate and deactivate the debounce function
>
I/O interface
>
>
In1 input
>
>
In1 function
>
Parameter
Factory setting
Teach-in
✓
Hold function
-
Measured value offset
-
Laser off
-
Off
-
Debouncing
The debounce function is only available in the
Advanced
user level.
When debouncing is activated, the input signal must be applied constantly to input In1
for 30 ms. The timing tolerances of the external teach-in functions take into account an
activation
or deactivation of the debounce function. It is not necessary to adjust the
timings.
>
I/O interface
>
>
In1 input
>
>
In1 function
>
>
Debouncing
>
Parameter
Factory setting
Debouncing?:
Yes, no
Yes
8.4.3.3
In1 active status
The input behavior can be selected as normally open (High Active, factory setting) or
normally closed (Low Active).
OPERATION VIA DISPLAY
8
8019642/18JN/2020-07-01 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD1000
51