5
Mounting
5.1
Mounting instructions
•
Observe the technical data.
•
Protect the sensor from direct sunlight.
•
To prevent condensation, avoid exposing the device to rapid changes in tempera‐
ture.
•
The mounting site has to be designed for the weight of the device.
•
To avoid inaccurate measurements when installing multiple devices: Make sure
that the laser light spot of one device is not in the visible range of another device.
•
Take into account the device warm-up time of 30 minutes. During the device
warm-up phase, the measured values are subject to an increased variance (tem‐
perature drift).
5.2
Mounting device
1.
Mount the device using the designated fixing holes,
.
2.
Make the electrical connection. Attach and tighten a voltage-free cable,
necting the device electrically", page 20
3.
Switch on the supply voltage.
✓
The green operating LED lights up.
The device needs around 10 seconds of initialization time before it is ready for
operation.
4.
Align the light spot so that the desired object is measured.
MOUNTING
5
8019642/18JN/2020-07-01 | SICK
O P E R A T I N G I N S T R U C T I O N S | OD1000
17